SUPPORT - MANUALS & DOWNLOADS
Instrument Inspector Targets

Rotated X-42, B-42 Targets

This manual helps to understand how the user can use any 3rd party software to scan X-42 or B-42 Targets in landscape position (rotated 90° clockwise).

Most handheld instruments  (i1Pro 1&2, eXact Scan, Techkon SpectraDens etc) will scan this way.

Some instruments require starting from white ( e.g. i1 Pro), some from the first patch  (eXact) - that's why in horizontal position there is a START/STOP WHITE area.

Typical scanning in a horizontal position will capture patches in following order:

first line (A): 36, 31, 25, 19, 13, 7, 1
second line (B): 37, 32, 26, 20, 14, 8, 2
...
last line (F): 42, 36, 30, 24, 18. 12, 6

 

Instrument Inspector patch recognition requires proper CMYK values for all 42 KEY patches:

ID  

C

M

Y

K

1

100.00

100.00

100.00

0.00

2

7.06

1.18

100.00

0.00

3

43.14

30.98

32.16

0.00

4

1.18

38.82

100.00

0.00

5

98.82

25.88

1.18

0.00

6

23.92

100.00

90.98

0.00

7

29.02

1.18

14.12

0.00

8

5.10

98.82

98.82

0.00

9

14.12

12.16

34.12

0.00

10

67.06

100.00

83.92

0.00

11

16.86

1.96

10.98

0.00

12

60.00

9.80

1.18

0.00

13

100.00

83.14

1.96

0.00

14

7.06

23.92

50.98

0.00

15

43.14

100.00

41.18

0.00

16

38.82

23.92

41.96

0.00

17

1.18

70.98

58.04

0.00

18

100.00

87.84

18.04

0.00

19

32.16

1.18

100.00

0.00

20

96.08

90.98

89.80

0.00

21

87.06

34.90

1.18

0.00

22

1.18

89.80

16.86

0.00

23

67.84

1.18

100.00

0.00

24

27.06

10.98

14.12

0.00

25

100.00

49.80

27.84

0.00

26

9.02

3.92

5.10

0.00

27

41.18

83.14

100.00

0.00

28

63.14

1.18

54.90

0.00

29

25.88

61.18

3.14

0.00

30

65.10

52.94

52.16

0.00

31

36.86

20.00

20.00

0.00

32

43.92

85.10

1.18

0.00

33

14.90

1.18

50.98

0.00

34

100.00

49.02

100.00

0.00

35

1.18

25.10

100.00

0.00

36

32.16

7.84

3.14

0.00

37

100.00

72.94

1.18

0.00

38

27.06

27.84

60.00

0.00

39

100.00

1.18

45.88

0.00

40

1.96

40.00

7.84

0.00

41

76.08

83.92

1.18

0.00

42

100.00

1.18

98.04

0.00

 

 

This list has to be reordered according to specific application requirements.

 

Additional Information

Instrument Inspector Targets all at least have 42 KEY patches.

The X-42 Target has 64 patches, the 22 additional patches are black or white control patches which our system will ignore if they are acceptable.


Control patches ignored by system (only for ChromaChecker Targets)

  • black patch definition C=0; M=0; Y=0; K=100
  • white patch definition: C=0,39;  M=0;  Y=0;  K=0
  • solid C, solid M, solid Y