ChromaChecker compared eXact to eXact2. The new instrument has many key improvements, and we recommend the upgrade.
CC Capture supports eXact2 today. We extensively tested the eXact2 instrument, one of which we did using QuickChecker's "Start Variation" feature. Our test started by checking the repeatability of the White calibration reading, and then we repeated a similar test using the black trap. The results of both tests are impressive - the main issues of the previous eXact are fixed perfectly. Also, eXact2 can measure shallow L values. In our test, the "black trap" returned an L* value as low as 0.15, which is a perfect result.


We took 50 measurements with irregular triggering intervals, making the test harder to pass. The final calculations were based on 1225 individual pairs. Comparison to the first generation shows extreme improvement. That addressed the weakest point.
X-Rite is making significant progress with new instrument generations. Like the i1Pro family, where each new generation offers a visibly better product, eXact2 addresses most of the issues of the first generation. We recommend an instrument upgrade for all demanding customers.
The current CC Capture version's Wi-Fi Connection is unstable. We are working on this issue. Please use a USB cable connection instead.
Additional information and Support Form is available for logged users.